- Connect BioAccoustic Simulator response switch to BioSteve (1/4 headphone jack)
- Connect BioSteve USB-C cable to Amtas
- Install USB-C to Serial adapter Driver (windows update or https://www.usconverters.com/usb-c-serial-rs232-adapter)
- Set COM port to COM5
-
- Open Device Manager
- Expand Ports COM and LPT
- Right Click USB to Serial (COM3) and select properties
- Under Port Settings Tab select Advanced
- Change COM Port Number to COM5
- Hit ok to save (may require reboot)
-
- Place Amtas Headphones over adapter plates on Bio Accoustic Simulator
- Download Bio Steve Script from https://assessor.ei1.com/download/biosteve20230405.exe and save to c:\programdata\Grason-Stadler\Amtas Flex\
- Create shortcut from Biosteve.exe to desktop (right click biosteve.exe and select send to desktop)
- Right click biosteve shortcut on the desktop and select properties. Set it to "Run Minimized"
- Create BioSteve shortcut for Amtas software
-
- From desktop – Right click GSI Amtas Flex shortcut and select copy (Person with headphones icon)
- Right click somewhere on the desktop and select paste to make a copy of the GSI Amtas Flex shortcut.
- Right click new shortcut and rename to “Run Calibration Test”
- Right click “Run Calibration Test” shortcut and select properties.
- In the target field paste the following
-
"C:\Program Files (x86)\Grason-Stadler\GSI Amtas Flex\GSI-AMTAS.exe" /pi Bio!Steve!!Male!1/1/1970!! /tf 500,1000,2000,3000,4000,6000,8000 /fx C:\ProgramData\Grason-Stadler\GSI-AMTAS /sv
- Hit ok to save
- Open GSI AMTAS Flex Config on the desktop (gear icon)
-
- Under Air Diagnostic
-
- Check Disable Masking
-
- Under Air Diagnostic
-